Stethoscope Troubleshooting
There is No Sound When Using the Stethoscope
- Try to move to a different, quieter room. Check the noise bar when the stethoscope is connected, and see if the behavior matches the desired volume levels.
- If the sound is still faulty or there is no sound, ask the customer to clean the stethoscope’s metal connectors using a dry or slightly moist wipe (not a wet one). Then, instruct them to insert the stethoscope securely into the Tyto device and try again.
- Check if the sound gauge responds to tapping lightly on the stethoscope.
- If there is no sound or no feedback from the sound gauge, ask the customer to factory reset their device:
- Menu (3 dots) → Settings → Advanced Settings → Reset Device
- If the sound from the stethoscope is still muffled / faulty, support should log in to the Clinician Dashboard, invite the user for a demo visit, select the Lungs / Heart exam and see if the issue persists.
- If there is muffled / no sound then contact TytoCare support.
Stethoscope Not Recognized
If the Heart/ Lung exam does not start after connecting the attachment and the screen is stuck on “Insert Stethoscope”:
- Ask the user to clean the stethoscope’s metal connectors with a dry cloth, insert the stethoscope, and try again.
- If the device still does not recognize ask the user to factory reset their device:
- Menu (3 dots) → Settings → Advanced Settings → Reset Device.
- After resetting, have them pair the device again and start an exam with the stethoscope attached.
- Further diagnosis if Issue persists:
- Ask the user whether the base unit fails to recognize a specific attachment (e.g., stethoscope or otoscope) or both.
- Instruct them to start a heart or lung exam with the stethoscope or an ear exam with the otoscope. This helps identify whether the issue is with a specific attachment or the Tyto base unit itself.